1. Starting the Service Software
1. Find the MicroscopeConfiguration.xml file in the folder below:
C:\\ProgramData\3DHISTECH\SideScanner\
2. Open it with some word processing software (for example,
Notepad).
3. Find the following line: <HardvarLayer Virtual=”False”
Type=Type_3”/>.
4. Modify the text with green background to Type_1 to activate service
mode.
Warning!
If more than one Instructor or Administrator operates at
the same time on the same virtual surface, they can erase, cancel, or
modify each others' savings.
5. Start the .exe file from the folder below:
C:\\Program Files\3DHISTECH\SlideScanner\MicroscopeService.exe
6. Enter the service password 'seristen'.
7. Click the Create new control profile button to create a new control
profile.
8. Set preset to 'P1000Prototype' and enter the serial number of the
instrument.
9. After using the software, reset Type_1 to Type_3.