5.2. To find the negative end position, navigate the unit in negative direction until the motor behaves as described in step 5. Press stop, and move
minus 200µm in positive direction. This is the negative end position, click on the side panel to set this value.
5.3. Set the Safety Z position using a Zeiss 40X immersion objective. Do the setting in the position, where the objective is above the center of the
slide support bar. This top position must be under the objective. Adjust the distance between the objective and the bar to 1mm with a gauge
plate, click on the side panel to set this value.
6. Validate the unit on the validation page.