P1000 SERVICE SOFTWARE MANUAL
14. SLIDE LOADING UNITS
64
3.4. Stage slide start Y
This is the position where the Gripper does not yet collide with the highest point of the Stage. This part is the tension spring. The bottom of
the Gripper must be +1 mm from the spring in Y. This can also be determined with a gauge plate. Click on the side panel to set this value.
1 mm