P1000 SERVICE SOFTWARE MANUAL
14. SLIDE LOADING UNITS
72
Here, you can do slide adjustments. If the slide in the magazine is not inserted until resistance and it sticks out too much, its sensor signs and the user
must insert the slides correctly. In case of a minor sticking out, there is no sign. To accord the preview image and the microscope image, the position
of the slide in the Gripper must be defined precisely. Therefore, it must be adjusted before taking a preview picture.
To do this adjustment, insert the marked double-wide slide into a magazine. Insert the adjustment slide to the magazine until resistance! Define its
position setting the roll-down menus and send it to the preview area clicking Slide to preview area.
5.1. Preview adjust pos X
This is the position, where the bisecting line of the gripped slide is aligned with the stop pin of the preview unit. Find and set this coordinate
by clicking on the side panel.