P1000 SERVICE SOFTWARE MANUAL
14. SLIDE LOADING UNITS
74
5.3. Preview adjust pos Z
On the Z-axis, the slide can be placed between the two pins so that it will not stall on them. Place the slide 1mm far from the back-pin setting
with a gauge plate. Click on the side panel to set this value.
5.4. Front adjust pos Z
This value sets the slide placement at the front pin. The slide must touch the pin, but it must not move in the Gripper. Click on the side
panel to set this value minus 1500µm.