The figure below shows the detection of a faulty adjustment.
The detected slides appear on the left side scale. Adjusting Sensor read offset slide bar, you can define the expected position of the slides
displayed as narrow lines on the right side. You must adjust the bar so that the sensor detects the inserted slides as it is in reality.
If calibration is not appropriate re-calibrate the Slide edge sensor according to the description
below. The sensor defines if there is a slide based on the intensity of reflected light. Open the
cover of the sensor.