English / 34-8720-2487-1 7
StandbyMode
The Standby mode is a temporary state that lasts a maximum of 10seconds and the
Bair Logo is displayed on the LCDscreen.
To enter Standby mode, press the Power button from Sleepmode.
To exit Standby mode, either: a) Press the Bair Logo for three seconds to enter
Diagnostic mode; or b) Wait 10seconds and the unit will go into Sleepmode.
SleepMode
During Sleep mode, the warming unit is idle and awaiting user activation. When in
Sleep mode the following eventsoccur:
• System timer remainsactive.
• The LCD goesblank.
• The Power indicator illuminatesamber.
To enter Sleep mode, press the Power button from either the Diagnostic mode or
Operationmode.
To exit Sleep mode, either: a) Press the Power to enter Standby mode; or, b) Press a
temperature button to enter Operationmode.
FaultConditions
The Model675warming unit’s software recognizes several nonhazardous fault
conditionsincluding:
• Sensorfailures.
• Microprocessorfailures.
• Softwarefailures.
• Keypadfailures.
• Power supplyfailure.
• Temperature out ofrange.
When a fault condition exists, the following eventsoccur:
• Blower turnsoff.
• Heater turnsoff.
• Audible alarmsounds.
• Fault wrench appears on LCDscreen.
• Fault code appears on LCDscreen.
• Up to 20fault codes are stored inmemory.
Silence Alarm/ResetUnit
• When the audible alarm sounds, press the Power button on the keypad to
silenceit.
• The warming unit must be unplugged before it can be used. See “Viewing the
Fault Code Log” and the “Fault Code Table” for more information about specific
faultcodes.
Over-TemperatureCondition
An independent electronic thermal cutoff safety circuit shuts the heater off to ensure
the hose end air remains below 56°C if an over-temperature condition is detected. An
over-temperature condition occurs when one of the following isdetected:
1. The independent thermal cutoff safety circuit senses the air exiting the unit
exceeds 53°C[±3°C].
2. The temperature exceeds 47°C for 30seconds.
3. The temperature is 3°C above set point for 3minutes.
When an over-temperature condition exists, the following eventsoccur:
• Over-temperature indicator and fault
condition wrench alternate displaying
in the LCDscreen.
• Audible alarmsounds.
• Blower and heater turnoff.
The warming unit must be unplugged before it can beused.
Under-TemperatureCondition
An under-temperature condition occurs when the warming unit detects the
temperature set point has not been reached or maintained during use. It occurs when
the temperature is less than 3°C below the selected temperature for 3minutes.
When an under-temperature condition exists, the following eventsoccur:
1. Under-temperature indicator and fault
condition wrench alternate displaying
in the LCDscreen.
2. Audible alarmsounds.
3. Blower and heater turnoff.
The warming unit must be unplugged before it can beused.
DiagnosticMode
The BairHugger Model675has a diagnostic mode that allows access to the service
and calibration utilities of the unit. You will need to access the diagnostic menu for
several service procedures described in thismanual.
To access Diagnostic mode, press and hold the Diagnostic Mode button (i.e. Bair
Logo) for three seconds. The Diagnostic Mode, Ambient, 32, 38, and Power buttons
are used to access Diagnostic mode, navigate through the Diagnostic Menu items, and
exit Diagnostic mode. The figure and accompanying table below show the front panel
button locations and functions used in Diagnosticmode.
DIAGNOSTICS MENU
FILTER TIMER
CALIBRATION
SYSTEM TEST
OT TEST
Scroll Up
Exit
Diagnostic
Mode
Select Menu
Item
Access Diagnostic
Mode/Main Menu
Scroll Down
Bair Hugger Button Functions in Diagnostic Mode
Button Action WhenPressed
Diagnostic
Mode
• Hold for 3seconds to access Diagnostic mode. The warming
unit must be in Standby mode or Fault mode to access
Diagnosticmode.
• When in any Diagnostic mode screen, press to go to the
DiagnosticMenu.
32 • Scroll down a Diagnostic Menuitem
38 • Scroll up a Diagnostic Menuitem
Ambient • Select a Diagnostic Menuitem.
Power
• Exit Diagnostic mode and return the warming unit to
Sleepmode.
The figure below illustrates the Diagnostic Mode menu structure and the associated
menuitems.
DIAGNOSTIC MENU
FILTER TIMER
CALIBRATION
SYSTEM TEST
OT TEST
SYSTEM INFO
FAULT LOG
BLOWER TIMER
43C
170W
340W
680W