8 3M™ Bair Hugger Model 675 Warming Unit – Service Manual
Diagnostic Mode menu item functionaldescriptions.
FILTER TIMER – Used to view the current filter timer and reset the filtertimer.
CALIBRATION – Used to adjust the temperature offset to calibrate the unit. The
warming unit uses the offset value to account for any temperature variation within the
warming unit and the heat loss through the hose to match the temperature displayed
on the LCD screen with the temperature of the air existing the hose. Calibration is
performed at43°C.
SYSTEM TEST – Used to view sensor, blower speed (RPM), and power information in
detail. The table below details the available tests to run andview.
Test* Action LCD Screen Description
43C
Turns the heater and
blower on and warms
the delivered air to the
43C set point for up to
30minutes
Display xx.x Calibrated hose endtemperature
PCS xx.x
Non-calibrated control
sensortemperature
PMS xx.x Monitor sensortemperature
RPMxxxx Blowerspeed
POWER xxx
Calculated power consumption
based upon heat usage and
dutycycle
170W
Turns the 170W heater
and blower on for
20seconds
340W
Turns the 340W heater
and blower on for
20seconds
PCS xx.x
Non-calibrated control
sensortemperature
PMS xx.x Monitor sensortemperature
RPM xxxx Blowerspeed
680W
Turns the 680W
heater and blower on
for 20seconds
*Depress the Ambient button to interrupt the test and return to the
SYSTEMTESTscreen
OT Test – Used to verify operation of the over-temperature monitoringsystem.
SYSTEM INFO – Used to view system timers and otherinformation.
LCD Screen Description
SYS TIMERhh:mm
Total time the unit has been plugged into a
powersource
OP TIMERhh:mm
Total accumulated time the unit has been running in
Ambient, 32°C, 38°, 43°modes
FIRMWAREx Current firmware version loaded in warmingunit
POWER SUPPLY xx.x
Monitor sensortemperature
Displays the 24volt power supply output involts
FAULT LOG – Displays the most recent 20faults of the warming unit on the LCD
screen in n FCx hh:mmformat.
LCD Screen* Description
1-FC-hh:mm n = Numerical order of fault code in thelog
: FC = Abbreviation for “faultcode”
n-FC-hh:mm x = Code for the faultcondition
:
hh:mm =
OP TIMER time the fault condition occurred,
(hh = hours; mm =minutes)
20-FC-hh:mm
*Depress the 32and 38buttons to scroll through thefaults
BLOWER TIMER – Displays the total time the blower has been active in hh:mm format
(hh = hours; mm =minutes).
Exiting the DiagnosticMode
Exit the Diagnostic mode by pressing the Power button or by waiting 60seconds with
noactivity.
ServiceProcedures
Installing and Replacing the PowerCord
ServiceFrequency
Asneeded
Tools/Equipment
• #2Phillipsscrewdriver
Method
1. Disconnect the warming unit from the grounded powersource.
2. If the warming unit is mounted to a rolling cart, remove the rolling cart
(See “Rolling Cart Removal and Installation”section).
3. Turn the warming unit upside down and support the warmingunit.
4. Remove the screws securing the two power cord clamps inplace.
Power Cord
Power Cord Clamp
Screw Location
Power Cord Clamp ScrewLocation
5. If replacing the power cord, pull the plug of the old cord out of the warming
unit and remove the power cord from the channel leading to the side of the
warmingunit.
6. Insert the plug of the new cord into the unit. (Use only 3Msupplied
powercords).
7. Route the power cord in the channel to the side of the warmingunit.
8. Replace the two power cord clamps to secure the power cord inplace.
9. If the warming unit was attached to a rolling cart, reinstall on the rolling cart
(See “Rolling Cart Removal and Installation”section).
10. Connect the warming unit to a properly grounded powersource.
Rolling Cart Removal andInstallation
ServiceFrequency
Asneeded
Tools/Equipment
• None
Method
1. Disconnect the warming unit from the grounded powersource.
2. Cart Removal: Gently pull on the front tab on the cart, lift the front and rotate
backward to detach the unit off thecart.
System
MODEL
675
CartRemoval