EasyManua.ls Logo

3M Bair Hugger 675 - Testing and Maintenance Procedures; Testing Over-Temperature Detection System; Viewing the Fault Code Log; General Maintenance and Storage

3M Bair Hugger 675
332 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 3M™ Bair Hugger Model 675 Warming Unit – Service Manual
If the temperatures do not match, adjust the warming unit’s
displayedtemperature:
a. Press the 32button to decrease the temperature or the 38button to
increase the temperature OFFSET value by0.1°C.
b. Wait for the DISPLAY temperature shown on the warming unit’s LCD
display to stabilize at 43.0°C ±0.1°C.
c. Continue to press the 32or 38button(s) until the DISPLAY temperature
in the LCD screen matches the temperature on the thermocouple
monitor as close aspossible.
d. Press the Ambient button to save the offsetsetting.
e. Press the Diagnostic Mode button to return to the Diagnostic Mode
main screen or the Power button to return to Sleepmode.
Note: If it is necessary to enter an offset value greater than ± 10°C, there may be
a defect in the warming unit’s sensor. Remove the unit from service until it can
berepaired.
Testing the Over-Temperature DetectionSystem
WARNING
Do not perform the over-temperature detection system test while the warming unit
is being used for warmingtherapy.
CAUTION
Connect each warming unit being tested to a separate powersource.
ServiceFrequency
Every 12months or 500hours ofuse.
Tools/Equipment
Model 22110Temperature TestUnit
Calibrated thermocouplemonitor
Method
1. Connect the warming unit to a grounded power source. The unit performs a
self-test and automatically proceeds to Standbymode.
Note: Ambient room temperature must be ≥ 20°C.
2. Connect the temperature test unit to the end of the warming unit hose by fully
inserting the hose end up to the hose card stopflange.
3. Access the Diagnostic mode (see Diagnostic Modesection).
4. Navigate to the OT TEST menu option then press the Ambient button. The
warming unit is now in OT TEST mode. In this mode, the warming unit performs
the followingtasks:
a. Blower runs for 3minutes.
b. Heater elements remainOFF.
c. LCD screen displays “PREPARING FOR OTTEST”
5. After 3minutes the warming unit performs the followingtasks:
a. Blower remainsON.
b. Heater elements turnON.
c. Over-temperature timeractivates.
d. LCD screen displays “TEST INPROGRESS”.
6. Wait for the warming unit to complete the over-temperature test (up to
210seconds). When the test is complete, the warming unit performs the
followingtasks:
a. Heater elements turnOFF.
b. Blower turnsOFF.
c. The audible alert sounds if the OT test completessuccessfully.
d. The LCD displays ”SENSOR OT DETECTED AFTER xx SECONDS
UNPLUG TO RESET” if the OT test completessuccessfully,
e. The LCD displays “SENSOR OT NOT DETECTED UNPLUG TO RESET” if
the OT test does not completesuccessfully.
Note: If the LCD displays “SENSOR NOT DETECTED UNPLUG TO
RESET” do not place the warming unit back into service and contact
3Mtechnicalsupport.
7. Verify the blower turnsoff.
8. Verify the maximum temperature measured by the temperature test unit at the
end of the hose is less than56°C.
Note: If it is 56°C or higher or the blower does not shut off, do not place the
warming unit back into service and contact 3Mtechnicalsupport.
9. Unplug the warming unit to return it to normaloperation.
10.Allow the unit to cool for at least five minutes before returning toservice.
Viewing the Fault CodeLog
ServiceFrequency
As often asnecessary.
Tools/Equipment
Nonerequired.
Method
1. Connect the warming unit to a grounded power source. The unit performs a
self-test and automatically proceeds to Sleepmode.
2. Access the Diagnostic mode (see Diagnostic Modesection).
3. Navigate to the FAULT LOG menu option then press the Ambient button. The
warming unit is now in FAULT LOG mode. In this mode, the warming unit
performs the followingtasks:
a. Blower remainsOFF.
b. Heater elements remainOFF.
c. The LCD displays faultcodes.
4. Press the 32and 38buttons to scroll through fault codes in the log. The
temperature management unit stores a maximum of 20faultcodes.
The table below summarizes the available fault codes and fault conditions. See the
“Troubleshooting” section for additionaldetails.
Fault Code FaultCondition
01 Control sensorfailure
02 Control sensor at least 47°C over-temperaturecondition
03 Monitor sensorfailure
04 Monitor or Control sensor over-temperaturecondition
50 Stuckkey
51 Blowerfailure
52 Watchdog timerfailure
70 Temperature out of rangehigh
71 Temperature out of rangelow
72 Rate of temperature increaseinsufficient
73 Control sensor and monitor sensormismatch
General Maintenance andStorage
ServiceFrequency
Asneeded.
CleaningInstructions
1. Disconnect the warming unit from the power source beforecleaning.
2. Cleaning should be performed in accordance with hospital practices for cleaning
OR equipment. After every use; wipe the warming unit, the outside of the warming
unit hose, and any other surfaces that may have been touched. Use a damp, soft
cloth and a hospital approved mild detergent, germicidal disposable wipes,
disinfecting towelettes, or antimicrobial spray.
3. Let air dry or dry with a separate clean softcloth.
NOTICE:
Do not use cleaning solutions with greater than 80% alcohol or solvents,
including acetone and thinner, to clean the warming unit or hose. Solvents may
damage the labels and other plasticparts.
Do not immerse the BairHugger warming unit in any liquid or subject it to any
sterilization process as this may damage the warming unit or itsparts.
Storage
Store all components in a cool, dry place when not inuse.

Table of Contents

Other manuals for 3M Bair Hugger 675

Related product manuals