10 3M™ Bair Hugger Model 675 Warming Unit – Service Manual
• If the temperatures do not match, adjust the warming unit’s
displayedtemperature:
a. Press the 32button to decrease the temperature or the 38button to
increase the temperature OFFSET value by0.1°C.
b. Wait for the DISPLAY temperature shown on the warming unit’s LCD
display to stabilize at 43.0°C ±0.1°C.
c. Continue to press the 32or 38button(s) until the DISPLAY temperature
in the LCD screen matches the temperature on the thermocouple
monitor as close aspossible.
d. Press the Ambient button to save the offsetsetting.
e. Press the Diagnostic Mode button to return to the Diagnostic Mode
main screen or the Power button to return to Sleepmode.
Note: If it is necessary to enter an offset value greater than ± 10°C, there may be
a defect in the warming unit’s sensor. Remove the unit from service until it can
berepaired.
Testing the Over-Temperature DetectionSystem
WARNING
Do not perform the over-temperature detection system test while the warming unit
is being used for warmingtherapy.
CAUTION
Connect each warming unit being tested to a separate powersource.
ServiceFrequency
Every 12months or 500hours ofuse.
Tools/Equipment
• Model 22110Temperature TestUnit
• Calibrated thermocouplemonitor
Method
1. Connect the warming unit to a grounded power source. The unit performs a
self-test and automatically proceeds to Standbymode.
Note: Ambient room temperature must be ≥ 20°C.
2. Connect the temperature test unit to the end of the warming unit hose by fully
inserting the hose end up to the hose card stopflange.
3. Access the Diagnostic mode (see Diagnostic Modesection).
4. Navigate to the OT TEST menu option then press the Ambient button. The
warming unit is now in OT TEST mode. In this mode, the warming unit performs
the followingtasks:
a. Blower runs for 3minutes.
b. Heater elements remainOFF.
c. LCD screen displays “PREPARING FOR OTTEST”
5. After 3minutes the warming unit performs the followingtasks:
a. Blower remainsON.
b. Heater elements turnON.
c. Over-temperature timeractivates.
d. LCD screen displays “TEST INPROGRESS”.
6. Wait for the warming unit to complete the over-temperature test (up to
210seconds). When the test is complete, the warming unit performs the
followingtasks:
a. Heater elements turnOFF.
b. Blower turnsOFF.
c. The audible alert sounds if the OT test completessuccessfully.
d. The LCD displays ”SENSOR OT DETECTED AFTER xx SECONDS
UNPLUG TO RESET” if the OT test completessuccessfully,
e. The LCD displays “SENSOR OT NOT DETECTED UNPLUG TO RESET” if
the OT test does not completesuccessfully.
Note: If the LCD displays “SENSOR NOT DETECTED UNPLUG TO
RESET” do not place the warming unit back into service and contact
3Mtechnicalsupport.
7. Verify the blower turnsoff.
8. Verify the maximum temperature measured by the temperature test unit at the
end of the hose is less than56°C.
Note: If it is 56°C or higher or the blower does not shut off, do not place the
warming unit back into service and contact 3Mtechnicalsupport.
9. Unplug the warming unit to return it to normaloperation.
10.Allow the unit to cool for at least five minutes before returning toservice.
Viewing the Fault CodeLog
ServiceFrequency
As often asnecessary.
Tools/Equipment
Nonerequired.
Method
1. Connect the warming unit to a grounded power source. The unit performs a
self-test and automatically proceeds to Sleepmode.
2. Access the Diagnostic mode (see Diagnostic Modesection).
3. Navigate to the FAULT LOG menu option then press the Ambient button. The
warming unit is now in FAULT LOG mode. In this mode, the warming unit
performs the followingtasks:
a. Blower remainsOFF.
b. Heater elements remainOFF.
c. The LCD displays faultcodes.
4. Press the 32and 38buttons to scroll through fault codes in the log. The
temperature management unit stores a maximum of 20faultcodes.
The table below summarizes the available fault codes and fault conditions. See the
“Troubleshooting” section for additionaldetails.
Fault Code FaultCondition
01 Control sensorfailure
02 Control sensor at least 47°C over-temperaturecondition
03 Monitor sensorfailure
04 Monitor or Control sensor over-temperaturecondition
50 Stuckkey
51 Blowerfailure
52 Watchdog timerfailure
70 Temperature out of rangehigh
71 Temperature out of rangelow
72 Rate of temperature increaseinsufficient
73 Control sensor and monitor sensormismatch
General Maintenance andStorage
ServiceFrequency
Asneeded.
CleaningInstructions
1. Disconnect the warming unit from the power source beforecleaning.
2. Cleaning should be performed in accordance with hospital practices for cleaning
OR equipment. After every use; wipe the warming unit, the outside of the warming
unit hose, and any other surfaces that may have been touched. Use a damp, soft
cloth and a hospital approved mild detergent, germicidal disposable wipes,
disinfecting towelettes, or antimicrobial spray.
3. Let air dry or dry with a separate clean softcloth.
NOTICE:
• Do not use cleaning solutions with greater than 80% alcohol or solvents,
including acetone and thinner, to clean the warming unit or hose. Solvents may
damage the labels and other plasticparts.
• Do not immerse the BairHugger warming unit in any liquid or subject it to any
sterilization process as this may damage the warming unit or itsparts.
Storage
Store all components in a cool, dry place when not inuse.