English / 34-8720-2487-1 11
Troubleshooting
The warming unit software routinely checks unusual operating conditions. If such a condition occurs, the system may trigger a fault condition and alarm. The warming unit can be
reset by disconnecting the warming unit from the main power source and allowing 30-60seconds for the system to reset. If the fault code condition reoccurs, disconnect again
and wait 5minutes for the system to reset. If the fault condition reoccurs upon connection to the main power source, the warming unit may have a hard-fault or permanent error.
If the fault does not reoccur upon connection to the main power source, the warming unit may have a cautionaryfault.
Fault Code Fault Condition Possible Cause Troubleshooting
01 Control sensor output
ishigh
1. Sensor boardfailure.
2. Control boardfailure.
3. Internal wire harness failure.
1. Check for moisture onsensor.
2. Perform the 43C test in the System Test option of the Diagnostic Menu.
Verify the end of hose temperature is 43°C ± 3°C using the Model
22110temperature test unit and calibrated thermocouplemeter.
02 Control sensor measures ≥
47°C for 30seconds
1. Control boardfailure.
2. AC board failure(TRIAC).
3. Sensor boardfailure.
4. Internal wire harness failure.
1. Perform Operating Temperaturecalibration.
2. Measure the hose end temperature using the Model 22110temperature
test unit and calibrated thermocouple meter while running in Ambient
mode. The measured temperature should be ±3°C of the ambient
roomtemperature.
03 Monitor sensor output
ishigh
1. Sensor boardfailure.
2. Control boardfailure.
3. Internal wire harness failure.
1. Check for moisture onsensor.
2. Perform the 43C test in the System Check option of the Diagnostic
Menu. Verify the end of hose temperature is 43°C ± 3°C using the Model
22110temperature test unit and calibrated thermocouplemeter.
04 Software detects an
over-temperature
circuitcutout
1. Control boardfailure.
2. Sensor boardfailure.
3. Internal wire harness failure.
1. Perform Operating Temperaturecalibration.
2. Measure the hose end temperature using the Model 22110temperature
test unit and calibrated thermocouple meter while running in Ambient
mode. The measured temperature should be ±3°C of the ambient
roomtemperature.
50 Software detects any button
pressed for 5seconds.
1. Useraction.
2. User interface circuit board failure.
1. Reset microprocessor by disconnecting the warming unit from the main
power source and allowing 30-60seconds for the system toreset.
51 The blower is running at ≤
2000RPM for 1minute, or
there is no output to the
control board for 3seconds
1. Blowerfailure.
2. Control boardfailure.
3. Internal wire harness failure.
1. Test blower using the System Info screen in Diagnosticmode.
52 Watchdog timer failed 1. Control board failure. 1. Reset microprocessor by disconnecting the warming unit from the main
power source and allowing 30-60seconds for the system toreset.
70 The control sensor measures
the temperature to be +3°C
greater than the set point for
3minutes.
1. Control boardfailure.
2. Air flow occlusion.
1. Check hose forobstructions.
2. Check the air intake on the bottom of the unit forobstructions.
3. Check for blockedfilter.
4. Test all 3heater wattages using the System Test screen in the
DiagnosticMenu.
71 The control sensor measures
the temperature to be -3°C
less than the set point for
3minutes.
1. Air flowocclusion.
2. Heaterfailure.
3. Control boardfailure.
4. Internal wire harness failure.
1. Check for hoseobstructions.
2. Check the air intake on the bottom of the unit forobstructions.
3. Check for blockedfilter.
4. Test all 3heater wattages using the System Test screen in Diagnosticmode.
72 The outlet temperature
does not increase at
expectedrate
1. Air flowocclusion.
2. Heaterfailure.
3. Control boardfailure.
4. Internal wire harness failure.
1. Check for hoseobstructions.
2. Check the air intake on the bottom of the unit forobstructions.
3. Check for blockedfilter.
4. Test all 3heater wattages using the System Test screen in Diagnosticmode.
73 Control and monitor sensors
differ by more than 3°C for
30seconds.
1. Sensor boardfailure.
2. Internal wire harnessfailure.
3. Control board failure.
1. Check for moisture onsensor.
2. Perform the 43C test in the System Check option of the Diagnostic
Menu. Verify the end of hose temperature is 43°C ± 3°C using the Model
22110temperature test unit and calibrated thermocouplemeter.
Technical Support and OrderPlacement
When you call for TechnicalSupport
We will need to know the serial number of your BairHugger warming unit when you
call us. The serial number label is located on the back of the warmingunit.
USA:
• Call 3MTechnical Support at800-733-7775
Outside of theUSA:
• Contact your local 3Mrepresentative
In-Warranty Repair andExchange
USA
• Call 3Mtechnical support at 800-733-7775if your warming unit requires
factoryservice.
• Provide the serial number located on the back of the warmingunit.
• Technical Support will give you a Return Authorization (RA) number. Use this RA
number on all correspondence. Your Technical Support representative will also
send a shipping carton to you at no charge, ifneeded.
Outside of theUSA
• Contact your local distributor concerning in-warranty repair and exchange.
Provide the serial number located on the back of the BairHugger warming unit
when youcall.