12 3M™ Bair Hugger Model 675 Warming Unit – Service Manual
AcronymDefinitions
EMI ElectromagneticInterference
EMC ElectromagneticCompatibility
ESD ElectrostaticDischarge
IV Intravenous
LED Light-emittingDiode
LCD Liquid CrystalDisplay
RF RadioFrequency
PCS Proximal ControlSensor
PMS Proximal MonitorSensor
RPM Revolutions PerMinute
PhysicalCharacteristics
Dimensions
14.5" high x 10.5" deep x 11"wide
(36.2cm high x 26.7cm deep x 28cmwide)
Weight 9.93lb (4.5kg)
Relative noise level 50dBA
Hose
Flexible
Length: 80" (203cm)
Weight: 1.51lb (0.686kg)
Diameter: 2" (5.08cm)
Unit Intake Filter MERV14*
Recommended Filter Change Every 12months or 500hours ofuse.
*See “Frequently Asked Questions on Filtration”section
Recommended operatingenvironment
Temperature 15°C-25°C(59°-77°F)
Humidity Up to 80% relative humidity(non-condensing)
Altitude Max 2,000m
Operatingtemperatures
Temperature control Electronicallycontrolled
Heat generated 2217BTU/hr (average), 650W(average)
Time to reach operating
temperature
2- 5minutes (dependent on blanketmodel)
Time required for the contact surface temperature to
heat up from 23± 2°C to 37°C (73± 2°F to99°F)
Storage/transport
temperature
-20°C to 45°C (-4°F to113°F)
Store all components in a cool, dry place when not inuse.
SafetySystem
Thermostat Independent electronic circuit; thermal cutoff shuts the heater
OFF to ensure hose end air remains below 56°C (53°C ±
3°Ctypical).
Alarm system Over-temperature (≤56°C, 53°C ± 3°C typical): red Over-Temp
indicator symbol will flash alternately with the “Wrench”
symbol, alarm sounds, heater and blower shut down, operating
indicator lights turn OFF, and control panel becomes
unresponsive. Remove fromservice.
Under-temperature (<3°C of selected temperature/setting
for 3minutes): blue Under-Temp indicator symbol will flash
alternately with the “Wrench” symbol, alarm sounds, heater
and blower shut down, operating indicator lights turn OFF, and
control panel becomes unresponsive. Remove fromservice.
Fault “Wrench” symbol displays and alarm sounds. Heater and blower
shut down, operating indicator light turns OFF and control
panel becomes unresponsive. Servicerequired.
Overcurrent
protection
Dual input fused lines. Fuses are to be accessed and/or
replaced only by servicepersonnel.
Alarm/FaultCodes
Alarm conditions Generalfault
Over-temperature
Under-temperature
Fault code indicator Yes on LCDscreen
Filter change indicator Yes: Triggered by 500hours ofuse
ElectricalCharacteristics
Heating element 1190WResistive
Heater control 3stageheater
flickerfree
Temperature settings Ambient
32° C (89.6°F)
38° C (100.4°F)
43° C (109.4°F)
Temperature accuracy +/- 3°C
Temperature display
resolution
0.1°C
Leakage current Meets UL 60601-1and IEC 60601-1requirements
Blower motor Non-adjustable, onespeed
Airflow: up to 44cfm or 20.77L/s
Operating speed: 4,700RPM
Power consumption Peak: 1255W
Average: 650W
Power cord 15ft., SJT, 3cond., 13A
15ft., SJT, 3cond., 15A
4.6m, HAR, 3cond., 10A
Device ratings 110-120VAC, 50/60Hz, 11.7A,or
220-240VAC, 50/60Hz, 6A,or
100VAC, 50/60Hz, 12A
Fuses Type: Fast acting ceramic fuses, 250VAC
AMPrating:
12A (110- 120VAC)
8A (220- 240VAC)
15A (100VAC)
Operating speed: F (FastActing)
Breakingcapacity:
15A, 12A: 750A @ 250Vac
15A, 12A: 10,000A @ 125Vac
8A: 200A @ 250Vac
8A: 10,000A @ 125Vac