12
Preparation
Starting
and
Priming
the
System
(Continued)
Prime
the
Attached
Equipment
Precaution:
The
unit
must
be
properly
primed;
running
the
pumps
and
heaters
with
the
device
dry
will
damage
them.
Remove
air
from
the
entire
system
including
attached
equipment:
'
1.
Run
in
Cooldown
mode
until
air
bubbles
disappear
from
the
external
tubing.
2.
Turn
the
Cardioplegia
on
in
Large
Tank
mode
and
run
until
all
air’bubbles
disappear
in
the
external
tubing.
Turn
the
Cardioplegia
off.
3.
Run
in
Rewarm
mode
to
remove
air
from
the
internal
warming
circuit.
If
air
bubbles
appear
in
the
external
tubing,
immediately
change
back
to
the
Cooldown
mode
until
the
air
bubbles
disap—
pear.
Then
alternate
between
Cooldown
and
Rewarm
modes
until
the
unit
will
operate
in
Rewarm
mode
without
external
air
bubbles.
Return
to
Standby
mode.
Check
the
Water
Level
Again.
Precaution:
The
water
level
in
the
large
tank
must
never
rise
above
the
maximum
level
as
water
may
damage
the
inside
of
the
unit;
this
water
level
should
not
drop
below
the
minimum
level
as
there
may
be
a
loss
of
system
performance
(see
maxi—
mum-minimum
label).
12
Test
the
Temperature
Probe
Modules
Depress
the
TEST
switch;
the
displays
must
read
25”
i
.4°C.
If
the
readings
fall
outside
this
range,
contact
Sarns
Service.
Release
the
test
switch
and
displays
should
show
probe
information
if
the
probes
are
connected.
Set
the
Alarm
Limits
Select
Probe
A,
B,
or
C;
the
probe
LED
will
light.
To
set
a
high
or
low
alarm
limit,
depress
the
appropriate
alarm
switch
[or
over
one
second
and
the
numbers
appearing
on
the
top
display
will
cycle
through
the
selection
range.
When
the
desired
number
appears,
release
the
alarm
switch
to
set
the
new
alarm
limit.
Check
the
Remote
Controls
Try
the
Maintain
Up
and
Down
switches.
Change
modes
and
check
that
modes
function
when
using
the
remote
unit
and
that
the
proper
LEDs
light
on
the
remote
and
main
panels.
If
the
unit
does
not
respond
properly
to
the
above
checks,
do
not
use
it;
contact
Sarns
Service,
your
local
3M
subsidiary
or
your
authorized
Sarns
distributor.
12