14
Operation
Resolving
Alarm
Conditions
When
an
alarm
condition
occurs,
first
try
the
suggested
procedure.
If
the
alarm
condition
continues,
do
not
use
the
equipment;
contact
Sarns
Service,
your
local
3M
subsidiary
or
your
authorized
Sarns
distributor.
Whenever
there
is
a
major
problem
with
the
system,
use
an
alternate
source
of
water
and
means
of
temperature
monitoring
until
the
situation
is
corrected.
Alarm
Condition
H20
>
42°C
Warning
Pump
Not
Primed
Warning
Low
Flow
to
Blanket
Warning
Service
Warning
Recurrence
of
Startup
Temperature
Display Module
panel
blank
due
to
problem
with
the
module
14
Procedure
Switch
to
Cooldown
mode
until
the
water
temperature
reaches
a
satisfactory
level:
the
warning
LED
will
go
out
and
the
buzzer
wilt
stop
when
the
temperature
drops
below
42°C.
Return
to
Rewarm
mode
if
desired.
if
this
warning
should
activate
with
a
minimal
load,
such
as
only
a
blanket
attached,
set
the
unit
in
Maintain
mode
at
about
40
degrees
until
the
additional
equipment
is
attached.
It
adequate
water,
try
selecting
any
mode
which
will
activate
the
main
pump
(Cooldown,
Maintain.
Rewarm
or
Defrost).
If
the
alarm
continues,
check
for
leaks.
Add
water
when
necessary.
Turn
the
power
Off
for
two
seconds
and
then
back
On.
Use
Cooldown.
Cardioplegia
and
Flewarm
to
complete
priming
the
system.
Reset
the
controls
and
continue
using
the
unit.
Check
for
leaks,
kinks,
or
blockage
in
the
tubing
for
the
blanket.
if
the
Water
Temperature
display
shows
zero
during
this
warning,
do
not
use
Maintain
and
Gradient
Rewarm
modes
as
temperature
will
increase
to
42°
C.
Monitor
the
patient
and
water
temperature
closely
to
protect
the
patient
and
to
assess
the
conditions
of
the
unit;
the
surgical
team
must
decide
whether
to
continue
using
the
SarnsTM
TCM
II
or
to
change
to
an
alternate
source
of
water,
Check
the
startup
conditions
are
correct.
Reset
controls;
continue
using
the
unit.
Unplug
the
remote
from
the
main.
Gradient
rewarming
may
be
continued
although
the
venous
blood
temperature
will
not
be
displayed.
14