7. Review and StartCycle
Review the selected cycle (Figure 46). Press Start to proceed and initiate the cycle. Select Back to change previously selected parameters of thecycle. Press Cancel to clear the cycle and
return to the mainscreen.
Open door
Sterilizer
Cycle: 55 ˚C
Aeration (hh:mm): 12:00
Load ID: 120413-1
Cartridge lot number: 2017-02AB
Operator ID: LMT
22.2
Menu
Menu
Start cycle:
˚
C
04/20/2015
05:27:42
64
6
2 3 4 5 61 7
Back
Back
Start
Start
Cancel
Cancel
98.0
kPa
Figure46.
Review and StartCycle
After the cycle is started, the GS Series sterilizer proceeds to the Preheat stage, the first of nine active stages of an EO sterilization cycle. Figure 47is the display screen of a cycle in the
Preheatstage.
Stop
Sterilizer
25.0
Menu
Menu
˚
C
04/20/2015
05:28:01
Cycle: Count:3355
˚
C
Show chart
84.5
kPa
15:30
Cycle time remaining (hh:mm):
Stage: Preheat
Open door
Figure47.
Preheat Stage DisplayScreen
DANGER: To reduce the risks associated with exposure to ethyleneoxide,
call 3M Health Care service personnel or authorized 3M service personnel immediately if there is a failure of the display or backlight and the 3M™Steri-Vac™ Sterilizer/Aerator GS Series
continues tooperate.
CAUTION: To reduce patient risks associated with exposure to potentially non-sterile devices or inadequate sterilizationprocedures,
call 3M Health Care service personnel or authorized 3M service personnel immediately if there is a failure of the display or backlight and the 3M™Steri-Vac™ Sterilizer/Aerator GS Series
continues tooperate.