41
12.2. Display ScreenIndications
Figure 48illustrates the important information presented on the display screen while the sterilization cycle isrunning. Table 4explains eachindication.
Sterilizer
55.0
Menu
Menu
˚
C
04/20/2015
01:41:59
Cycle: Count:3355
˚
C
Show chart
98.0
kPa
Stop
Stage: Flushing
C604 - Power interruption
Details...
04/20/2015 06:50:57
EO
E109 -
Details...
04/20/2015 01:38:08
Open door
Open unload
door
0:00
Cycle time remaining (hh:mm):
1:00
Gas expose time elapsed (hh:mm):
12:00
Aeration time elapsed (hh:mm):
Cycle failure
Error level recovery
Cycle complete
RH
50%
8
8
6
9
1 2 3 4 5 7
10
11
12
13 14 15 16 17
18
Figure48.
Display ScreenIndications
Indication
Number IndicationDescription
1 EO Icon.
Indication ethylene oxide (EO) is in thechamber.
2 Shows current chambertemperature.
3 Shows current chamberpressure.
4 Shows current chamber percent relative humidity (%RH). %RH is only measured and displayed during the Conditioningstage.
5 Indication that Maintenance is due and Service should benotified.
6 An orange, downward arrow symbol will appear on the display screen task bar to remind the user that a pending software update is available forinstallation.
7 Shows current date andtime.
8 Error Message. Indicates a fault has occurred. See Chapter 15: Cautions, Error Messages and Troubleshooting for specific errors andcorrections.
9 Indication that a used 3M™ Steri-Gas™ EO Gas Cartridge is in thechamber.
10 Indication that an unused 3M™ Steri-Gas™ EO Gas Cartridge is in thechamber.
11 Counter. This is the sterilizing cycle counter that indicates the cumulative number of all cycles started, including those cycles in which a faultoccurred.
12
Cycle stage icons. While a sterilization cycle is running, the icons on the bottom of the display screen indicate the current stage of the cycle (indicated by a white border around the
icon). Each icon and the corresponding sterilization cycle stage are described in the section Overview of the GS Series EO Sterilization Cycle.
13 Menu button is used to access GS Series sterilizer options (e.g. cycle reports, setupvariables)
14
Open Door button (the word “load “is only used for double door or pass through sterilizer models). During the sterilization process, the Operator cannot open the door; therefore, the
button is coloredgray.
15
Open Unload Door button is used on double door pass through GS Series sterilizer models. During the sterilization process, the Operator cannot open the door; therefore, the button is
coloredgrey.
16 Show Chart button displays a running graph of the current cycle inprocess.
17 Stop button cancels the cycle in progress. There are times during the sterilization process when the Operator cannot cancel the GS Series sterilizer; therefore, the button is coloredgrey.
18 Cycle Complete signifies the Operator programmed cycle is successfully complete.
Table 4. Explanations of Display ScreenIndications