53
Steri-Vac
TM
Sterilizer GS5-1D
Serial number:
Software version:
Facility:
Sterilizer ID:
Start time:
Cycle count:
Cycle name:
Operator ID:
Load ID:
EO cartridge lot:
AA121212
1.0.1
City Hospital
Machine 2
04/20/2015 05:27:42
33
55° C
LMT
120413-1
2017-02AB
Figure61.
Verifying Cycle Report –Header
WARNING: To reduce the risks associated with fire andexplosion:
Inspect the display screen and cycle reports for errors and listen for audible notifications (if enabled). Always take action for errors as indicated in thismanual.
CAUTION: To reduce patient risks associated with exposure to potentially non-sterile devices or inadequate sterilizationprocedures:
Inspect the display screen and cycle reports for errors and listen for audible notifications (if enabled). Always take action for errors as indicated in thismanual.
Always inspect cycle reports (printout or electronic file) to ensure the Operator’s programmed parameters or the device manufacturer’s instructions for use (IFU)matches:
• %RH at the End ofConditioning,
• Temperature at the End ofConditioning,
• Actual Gas ExposureTime.
Always follow device manufacturer’s instructions for use (IFU), including device cleaning, drying, packaging, sterilization, andaeration.
Cycle complete
No errors
No cautions
End time: 04/20/2015 21:03:12
59%
55.0° C
1:00 hh:mm
12:00 hh:mm
15:30 hh:mm
RH end of
conditioning:
Temp end of
conditioning:
Actual expose time:
Actual aeration time:
Total cycle time:
Reviewed by & date:
15 21:03:12
Final StatusConfirmation
Temperature End ofConditioning
%RH End ofConditioning
Actual EO Gas ExposureTime
Actual AerationTime
Figure62.
Verifying Cycle Report -
F
ooterInformation