80 81
16. Cautions, Error Messages, andTroubleshooting
DANGER: To reduce the risks associated with exposure to ethylene oxide:
Never use force to access the inside of the sterilization chamber.
Inspect display screen and cycle reports for errors and listen for audible notifications (if enabled). Always take action for errors as indicated in this manual.
Call 3M Health Care service personnel or authorized 3M service personnel immediately if there is a failure of the display or backlight and the sterilizer continues to operate.
WARNING: To reduce the risks associated with fire and explosion,
inspect the display screen and cycle reports for errors and listen for audible notifications (if enabled). Always take action for errors as indicated in this manual.
CAUTION: To reduce patient risks associated with exposure to potentially non‑sterile devices or inadequate sterilization procedures.
inspect the display screen and cycle reports for errors and listen for audible notifications (if enabled). Always take action for errors as indicated in thismanual.
The 3M™Steri‑Vac™ Sterilizer/Aerator GSX Series is designed with embedded software that automatically controls, monitors and regulates the mechanical functions of the GSX Series sterilizer
and the sterilizationprocess.
16.1. CautionMessages
If the 3M™Steri‑Vac™ Sterilizer/Aerator GSX Series detects a fault condition that does not adversely affect the performance of the sterilization cycle, a caution message will appear on both the
display screen and cycle report to notify the Operator (Figure95).
The cycle can complete successfully when a caution message is asserted. Document all caution messages per facilityprocedures.
04/20/2015 08:29:56
Error level: C
Stage : EO injection
Info: SystemModeStateMachine
C604-Power interruption
Hide
Clear
Letter ‘C’ indicates message
is a caution. Document per
facility procedures.
Figure95.
Example of a CautionMessage