16.3. Error Levels and CorrectiveActions
Error
Level Description Example Error Corrective Action
L1
Errors that occur before ethylene oxide (EO) gas is in the
chamber and does not require immediate Service attention.
Chamber too hot to start a
newcycle
Timeout attempting to
reachtarget
%RH setpoint
Operator can clear error and accessload.
Wait until Open Door or Open Load Door button isblue.
Press OpenDoor.
Press Clear to clear theerror.
Document per facilityprocedures.
Restart cycle per facility procedures.
L2
Errors that occur before EO gas is in the chamber and
require Service attention. Operator can open load door
before Service arrives.
Cooling fan failure before
EOInjection
12V/24V power supply failure
before EO Injection
Operator can accessload.
Wait until Open Door or Open Load Door button isblue.
Press OpenDoor.
Call Service. Service must clear theerror.
Document per facilityprocedures.
Restart cycle per facility procedures.
L3
Errors that occur before EO gas is in chamber and requires
Service attention before Operator can open load door.
Puncture pin not detected in
home position (pin deployed) in
idlestate
Service aborted cycle
CallService.
Service must access load and clear theerror.
Document per facilityprocedures.
Restart cycle per facility procedures.
L4
Errors that occur after EO gas is in the chamber. Pressure
and temperature sensors can return system to a safe state.
Supervisor or individual with higher access must clear error.
System will then return to main screen where door can be
opened.
Monitoring sensor failures after
EO Injection
Supervisor or Service must clear theerror.
Wait until the Stop button isred.
Supervisor or Service press Stop Cycle and enter PIN. Supervisor
or Service press Clear and enter PIN to clear theerror.
Document per facility procedures.
L5
Errors that occur after EO is in the chamber. Pressure and
temperature sensors can return system to a safe state;
however, Service is required to open load door.
Leak detected in EO
Exposestage
Puncture pin not detected as
retracted in Gas Injection stage
CallService.
Service must access load and clearerror.
Document per facilityprocedures.
L6
Errors that occur after EO is in the chamber. Pressure
sensors can return system to a safe state but temperature
sensors cannot. Service attention is required.
Chamber over temperature and
under temperature failures after
EO Injection
CallService.
Service must access load and clear error. Document per
facilityprocedures.
L7
Errors that will require 3M Service. EO may or may not be
present in the chamber. Errors that occur during an Error
Recovery stage will be elevated to L7. Service attention is
required.
Failure in controller pressure
sensor after EOInjection
Exhaust line blockage detected
CallService.
Service must clear error. Document per facilityprocedures.
A1 Not Used Not applicable Not applicable
A2
Errors in temperature control that occur during programmed
aeration. Service attention is required.
Chamber over temperature
or under temperature during
aeration
CallService.
Service must access load and clearerror.
Document per facility procedures.
A3
Errors that occur during programmed aeration and require
Service attention.
Loss of compressed air
duringaeration
Power supply failures during
aeration
CallService.
Service must access load and clearerror.
Document per facilityprocedures.
Table8. Descriptions for Error Levels and CorrectiveActions