User Manual FOX 514
September 2021
File: User Manual_FOX 514_rev3 30
6.2 Starting the Device
Switch on the device by pressing the plus and minus keys simultaneously for at least two
seconds. The green starting LED will begin to flash. After about 30 seconds, the startup
screen will appear.
The FOX 514 Laser executes its system check automatically upon startup.
Once the system check has been completed, you have to enter the password (“0000”). See
section 6.4.3.3 to modify the password. After inserting the correct password, you will be
automatically redirected.
A T T E N T I O N
The FOX 514 should not be operated for more than five hours at a time. Restart the
laser after five hours at the latest so that it can carry out the system check during the
start routine. This ensures that undetected error cannot occur.