User Manual FOX 514
September 2021
File: User Manual_FOX 514_rev3 42
Output power measurements
• Check the set power with an external power meter (permissible tolerance + 20%)
A T T E N T I O N
If one or more safety-related points are objected to after the safety-related inspection
(STK), the device may no longer be operated.
8.3 Care by the user
The following maintenance instructions can be carried out by the user. These serve to make
your work easier. The system must be disconnected from the mains for cleaning. Always
use a damp, but never wet, soft cloth for cleaning and disinfecting.
For cleaning and disinfection, the power supply and the applicator must be disconnected. In
addition, the red dust protection cap should be insertet into the fiber connector. First clean
the device with clear water and neutral detergent to remove coarse and visible
contamination. Make sure that no moisture penetrates into the device.
Furthermore, wipe disinfection is possible. When choosing the disinfectant, pay attention to
the following:
• according to the manufacturer of the disinfectant, the disinfectant should be suitable
for non-invasive medical devices
• according to the manufacturer of the disinfectant, the disinfectant should be suitable
for wipe disinfection of surfaces
• the disinfectant should be based on alcohol and/or quaternary compounds
• the disinfectant should be suitable for lacquers
• aldehyde-free disinfectant (recommended)
When wiping disinfection, the manufacturers instruction for the disinfectant must be
respected. Following possible disinfectants that meet the above mentionted requirements
are listed:
Mikrobac forte, Bacillol 30 Foam
Acryl-des, antifect AF (N)
Incidin Foam, Incidin Pro
Neoform MED AF, neoform MED rapid
*This list is not exhaustive.
A T T E N T I O N
When cleaning the device, it must always be switched off and disconnected from the
mains. Wet wipes should be avoided in any case. The penetration of water or
disinfectant can lead to defects.