Output power measurements
Check the set power with an external power meter (permissible tolerance + 20%)
A T T E N T I O N
If one or more safety-related points are objected to after the safety-related inspection
(STK), the device may no longer be operated.
8.3 Care by the user
The following maintenance instructions can be carried out by the user. These serve to make
your work easier. The system must be disconnected from the mains for cleaning. Always
use a damp, but never wet, soft cloth for cleaning and disinfecting.
For cleaning and disinfection, the power supply and the applicator must be disconnected. In
addition, the red dust protection cap should be insertet into the fiber connector. First clean
the device with clear water and neutral detergent to remove coarse and visible
contamination. Make sure that no moisture penetrates into the device.
Furthermore, wipe disinfection is possible. When choosing the disinfectant, pay attention to
the following:
according to the manufacturer of the disinfectant, the disinfectant should be suitable
for non-invasive medical devices
according to the manufacturer of the disinfectant, the disinfectant should be suitable
for wipe disinfection of surfaces
the disinfectant should be based on alcohol and/or quaternary compounds
the disinfectant should be suitable for lacquers
aldehyde-free disinfectant (recommended)
When wiping disinfection, the manufacturers instruction for the disinfectant must be
respected. Following possible disinfectants that meet the above mentionted requirements
are listed:
Manufacturer
Possible disinfectants*
BODE Chemie GmbH
Mikrobac forte, Bacillol 30 Foam
ANTISEPTICA
Arcylan, Biguacid Liquid
Schülke & Mayr GmbH
Acryl-des, antifect AF (N)
Dr. Schuhmacher GmbH
CLEANISEPT; Descosept AF
Ecolab
Incidin Foam, Incidin Pro
Dr. Weigert
Neoform MED AF, neoform MED rapid
*This list is not exhaustive.
A T T E N T I O N
When cleaning the device, it must always be switched off and disconnected from the
mains. Wet wipes should be avoided in any case. The penetration of water or
disinfectant can lead to defects.