User Manual FOX 514
September 2021
File: User Manual_FOX 514_rev3 44
8.5.2 Troubleshooting
The following errors cannot be identified by the system self check and have to be checked
by the user:
Connect the laser to the mains and
restart it.
As a standard, the laser enters
the power save mode after 5
min of inactivity. Touch the
display or a blue arrow-key to
activate the device.
Laser is in Stand-By mode
The aiming beam is only visible
when the laser is in READY
mode. Switch the laser to READY
mode
Aiming beam brightness
set too low
Increase the parameter for
aiming beam brightness
Fiber is not properly inserted
Twist the fiber connector inside the
socket with a light movement in
wards
Replace the broken fiber with a
new one
Aiming beam - diode defective
Contact your A.R.C. Laser
representative
No laser beam but aiming
beam visible
Foot switch non plugged in
Check if foot switch is correctly
plugged in
Contact your A.R.C. Laser
representative
No aiming beam and no laser
beam visible
Fiber optic is not connected
to laser or not correctly
plugged in
Check the correct fitting of the
fiber or hand piece
Replace the broken fiber with a
new one