ABB Advance Optima magnos206 - Configuring Status Signals 7

ABB Advance Optima magnos206
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
42/24-10 EN Rev. 9 Chapter 7: Gas Analyzer Configuration 7-C-11
Configuring Status Signals
Menu Path MENU Configure System Status signals
Function The status signal configuration is set at the factory per the customer’s order.
Generally this configuration does not have to be changed in the field.
Selection The following signals are available:
Individual status signals, i.e. Error, Maintenance Request and Maintenance Mode
Overall status signal
If the status signal configuration is changed from “overall status signal” to “indi-
vidual status signal”, the digital outputs DO2 and DO3 of the standard function
block application “status signals/ externally controlled calibration” which possibly
are assigned to limit value signals will be overwritten with single status signals.
For further information on status signals, see the “System Status: Status Signals”
section, page 10-4.