42/24-10 EN Rev. 9 Chapter 8: Gas Analyzer Calibration 8-A-11
Calibration Methods, continued
Overview The following table summarizes the various calibration methods.
Quantity Calibration
SC MR Method To configure ... Calibrate ... Calibration affects ...
1 1 Test Gas/
Single
• Zero
• Span
individually for each sample
component and range
Measurement range
≥ 1 > 1 Test Gas/
Common
Measurement ranges for
zero and span calibration
• Zero in one measurement
range
• Span in another range
for each sample component
All sample component
measurement ranges
> 1 ≥ 1 Substitute
Gas
Components and
measurement ranges for
zero and span calibration
• Zero in one component
measurement range
• Span in one range for
another component
for each detector
All detector components
and measurement ranges
SC = Sample and Substitute Gas Components
MR = Measurement Ranges per Component
Setting the
Calibration Method
(see Figure 8-A-7)
The calibration method can be set separately for manual, automatic and externally
controlled calibration.
For common and substitute gas calibration the sample ranges for start- and end-
point calibration of all three types of calibration control are adjusted jointly.
For substitute gas calibration the zero and span calibration components should
also be set.
Figure 8-A-7
Setting the
Calibration Method
Substitute
gas
Zero range Span range
Common
calibration
Single
calibration
Zero range Span range
Zero
component
Span
component
Test gas
Component
Calibration
method