42/24-10 EN Rev. 9 Chapter 8: Gas Analyzer Calibration 8-A-11
Calibration Methods, continued
Overview The following table summarizes the various calibration methods.
Quantity Calibration
SC MR Method To configure ... Calibrate ... Calibration affects ...
1 1 Test Gas/
Single
Zero
Span
individually for each sample
component and range
Measurement range
1 > 1 Test Gas/
Common
Measurement ranges for
zero and span calibration
Zero in one measurement
range
Span in another range
for each sample component
All sample component
measurement ranges
> 1 1 Substitute
Gas
Components and
measurement ranges for
zero and span calibration
Zero in one component
measurement range
Span in one range for
another component
for each detector
All detector components
and measurement ranges
SC = Sample and Substitute Gas Components
MR = Measurement Ranges per Component
Setting the
Calibration Method
(see Figure 8-A-7)
The calibration method can be set separately for manual, automatic and externally
controlled calibration.
For common and substitute gas calibration the sample ranges for start- and end-
point calibration of all three types of calibration control are adjusted jointly.
For substitute gas calibration the zero and span calibration components should
also be set.
Figure 8-A-7
Setting the
Calibration Method
Substitute
gas
Zero range Span range
Common
calibration
Single
calibration
Zero range Span range
Zero
component
Span
component
Test gas
Component
Calibration
method