42/24-10 EN Rev. 9 Chapter 8: Gas Analyzer Calibration 8-C-9
Magnos206: Single-Point Calibration, continued
Calibration data for manual calibration
Calibration method Common single-point calibration
Calibration measurement range 0-25 Vol.-% O
2
Test gas concentration 20.96 Vol.-% O
2
Calibration data for automatic calibration
Calibration method Common calibration (test gas)
Single zero calibration always
Single span calibration never
Common zero and span calibration never
Test gas concentration zero gas 20.96 Vol.-% O
2
Calibration Data for an
Analyzer Module with
1 Sample Component
(Example:
Test Gas = Air)
Test gas concentration span gas n.a.
Calibration data for manual calibration
Calibration method Substitute gas calibration
Zero component O
2
in N
2
Zero range 0-25 Vol.-% O
2
Span component n.a.
Span range n.a.
Test gas concentration zero gas 20.96 Vol.-% O
2
Test gas concentration span gas n.a.
Calibration data for automatic calibration
Calibration method Substitute gas calibration
1)
Single zero calibration always
Single span calibration never
Common zero and span calibration never
Test gas concentration zero gas 20.96 Vol.-% O
2
Calibration Data for an
Analyzer Module with
> 1 Sample Component
(Example:
Test Gas = Air)
Test gas concentration span gas n.a.
1) Settings of sample components and measurement ranges for zero and span
calibration are taken over from the calibration data for manual calibration.