10-18 Chapter 10: Status Messages, Troubleshooting 42/24-10 EN Rev. 9
Gas Analyzer Problems
Blinking Measurement Value Readout
Measured signal
violates
measurement range
limits
Note:
Measurement value > +130 % MRS or measurement value < –100 % MRS
Additionally, status messages 344 or 345 are generated.
Blinking --E-- in Measurement Value Readout
Problem in measured
signal processing
View status messages.
Identify cause and repair.
Blinking --E-- in mA Value Readout
Problem in output
current circuit
Identify cause (e.g. line break) and repair.
Flow Problem
External gas lines or
filters dirty, plugged
or leaking
Disconnect the gas analyzer from the gas preparation system.
Blow out the gas lines with compressed air or clear them mechanically.
Change the filter elements and packings.
Check gas line seal integrity.
Gas analyzer gas
paths crimped or
leaking
Disconnect the gas analyzer from the gas preparation system.
Check the analyzer module gas lines and the gas module lines for crimping or
loose connections.
Check the integrity of the analyzer module gas paths and (if applicable) of the
lines to the gas module (see page 9-2 for instructions).
Continued on next page