SHB_U14 Chapter 6: Component Interchange 6-7
Change sample cell
Removing/installing
the sample cell
To remove the sample cell, proceed as follows:
Step Action
1 Carry out "Analyzer module removal" steps 1-7.
2 Remove the infrared detector wiring.
3 If applicable, remove the calibration unit wiring.
4 Loosen and bend back the clamp on the base support.
5 Remove the hoses from the sample cell.
6 If applicable, remove the 200-mm cell attachment O-ring
7 Slide the entire optical unit out of the base support.
8 Depending on the option involved, disassemble the optic unit according
to the figure.
9 Reverse the above steps to install.
Figure 6-7
Removing without
calibration cell
Continued on the following page