SHB_U14 Chapter 6: Component Interchange 6-9
Change calibration cell and calibration unit
Removing the
calibration cell
To remove the calibration cell, proceed as follows:
Step Action
1 Carry out all steps listed under "Sample cell removal".
Installing the
calibration cell
Caution!
When installing, make sure to place the calibration cell window which is filled
with test gas in the proper side.
The left and right beam paths are mirror images of each other.
Figure 6-9
Calibration unit with
calibration cell