6-10 Chapter 6: Component Interchange SHB_U14
Change infrared detector
Removing/installing
the infrared detector
To remove the infrared detector, proceed as follows:
Step Action
1 Carry out "Analyzer module removal" steps 1-7.
2 Remove all wiring connected to the optical components.
3 Loosen and bend back the clamp on the base support.
4 Remove the hoses from the sample cell.
5 If applicable, remove the 200-mm cell attachment O-ring
6 Pull the entire optical unit out of the base support.
7 Remove the infrared detector per Figure 5-8 and 5-10.
8 Reverse the above steps to install
Make sure the proper end disk (bright or dark) is installed behind the
infrared detector.
Figure 6-10
Removing the
infrared detector