7-4 Chapter 7: Configuration SHB_U14
Software configuration with TCT
TCT operating
environment
Test and calibration software (Windows version)
Accessories
PC serial port interface cable
PC with Windows 95, Windows NT, etc.
When should a TCT
configuration be
performed?
• When setting up a new component
• When setting up a new measurement range
• When adding or removing the calibration unit
• When changing the user text
• When changing the type of communication
General instructions
for TCT software
setup
The "Configure" menu item has the following elements
• General data
• Detectors
• Gas Modules
“General data” contains information such as serial number, network class, transfer
rates, etc.
“Detectors“ contains all detectors configured in the module, as well as all
measurement detectors and auxiliary detectors (temperature, pressure, flow).
For each detector the configured components and associated measurement
ranges are shown together with the applicable correction factors for detectors,
components and measurement ranges.
The calibration cells are configured in “Gas modules”.
Continued on next page