EasyManua.ls Logo

ABB Advance Optima Uras 14 - Initial Calibration of Uras 14 IR Detectors

ABB Advance Optima Uras 14
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SHB_U14 Chapter 8: Calibration
8-9
Initial calibration of Uras 14 IR detectors
Accessories Zero gas
Span gas for large measurement range
Gas mix pump as needed
PC with TCT
When should an
initial calibration be
performed?
Change in configuration
Set up of a new IR detector
Change in large measurement range
Installation of sample cell with different length
Retrofit of optical filters
Retrofit of filter cells
Retrofit of sample/filter cells
Retrofit of calibration cells
Change in emitter performance
Change in amplification
Change in analyzer module installation location (vertical/horizontal)
Change of temperature correction to temperature control and vice versa
Change of components
IR detector
Emitter insert
Note An initial calibration of each IR detector installed is performed at the factory. The
setting is made with dry gases.
Perform the initial calibration only with the hood cover removed.
The analyzer module should be in a location as free of vibration as possible.
Prior to performing an initial calibration, perform an optical and phase alignment.
The initial calibration should always be performed with test gases having a known
concentration.
Span gas concentration:
70 to 110 % of the large measurement range
Only perform the initial calibration approx. 60 minutes after the end of the warm-
up phase. (warm-up phase approx. 60 min)
Monitor using the Diagnostics/controller parameter variables menu, or TCT device
test.
Linearization of initially calibrated IR detectors is automatic and is not
measurement range-dependent.
Perform postlinearization to check.
(Exceptions: This does not apply in case of preset polynomial linearization)
Prior calibrations are cancelled by an initial calibration.
Any linearity placed in effect will remain unchanged. (e.g. linearity parameters will
be retained)
If, for example, no postlinearization can be performed (no mixture pump on site)
and an IR detector was changed, both postlinearization correction factors must be
set to 1 with TCT. This corresponds to the default setting
After an initial calibration the calibration cell component must be redetermined.
When installing a sample cell with a different length the new length must be
entered in the TCT "Detector correction parameter" menu.
Reason: Linearity would not be correct otherwise.
Continued on next page

Table of Contents

Related product manuals