8-26 Chapter 8: Calibration
SHB_U14
Flow detector initial calibration
When should an
initial calibration be
performed?
After changing the flow sensor
In case of a fault: Pump off Flow detector does not indicate "zero"
Accessories PC with serial port
Advance Optima - TCT test and calibration software
Serial port connecting cable, item no. 0743091
Test Gas
Flow measurement unit
Needle valve
Preparation Check gas path seal integrity
Establish electrical connections
The analyzer should be installed at the site at which it is to be used and should
be protected against vibration
Set the test gas flows between 30 and 60 l/h and maintain stable +/-5 l/h.
Connect the PC to the analyzer module
Start the calibration and test software
Test Test all detectors (the measured values should be stable and regularly updated)
Status message 0 for all detectors
Continued on next page