5-8 Chapter 5: Testing SHB_U14
Sample cell
Visual check
Since the sample gas flows throughout the sample cell’s sample chamber, the
optical properties of its inner surface can change.
Disassemble the cell for a visual check.
Hold the cell up to a light and look through it.
Both chambers must have equal and uniform surfaces with a bright metallic sheen
and no dark spots.
The windows must be clear and clean.
When? Offset too high
Zero drift
Loss of sensitivity
Causes of staining
and contamination
Sample chamber
Process (sample gas composition)
Incomplete or improper sample gas preparation
Inadequate maintenance
Reference chamber/extension chambers
Loose chambers
Other measures Clean the sample cell sample chamber
Replace the sample cell
Check the tubing and replace if necessary.
Eliminate the cause