…9 Specification
… Environmental and process conditions
Upstream and downstream pipe conditions
(octagonal bore DN40 to 200)
5 x pipe diameter
Flow direction
0 x pipe diameter
Upstream and downstream pipe conditions
(full bore DN250 to 2400)
5 x pipe diameter
Flow direction
2 x pipe diameter
Pressure loss
Flow rate Pressure loss in bar (psi)
Q
Q
Negligible for full bore sensors
Potable water approvals for wetted parts
WRAS
ACS
DVGW
NSF
AZ/NZ 4020
Metrological certifications (pending)
• OIML R49 for accuracy class 1 and 2 and environmental
class T50 0.1 to 50 °C (32.18 to 122 °F)
• MID MI-001
• NMI R49 and NMI 10 for Australia
• PAC for China
DS/FET400-EN