ENDURA AZ20 & AZ30 | OXYGEN ANALYZER | SI/AZ20/AZ30-EN REV. A
5 Zirconia Probe as part of the
safety function system
Failure rates
Notes:
1 Auto-diagnosed failures are assumed to be dealt within 48
hours. Undiagnosed failures are assumed to be revealed
subject to a proof-test interval of one year.
2 The equipment is in virtually continuous use such that failure
rates (based on 8760 hours per year) are applicable.
6 Functional safety compliance -
prerequisites to operate the device
The analog signal of the transmitter can be considered safe
after 2 hours (warm up time), or after the heater output
stabilizes (see signal view).
A dangerous error is an error during which the output of the
transmitter no longer responds to the input signal or deviates
by more than 2 %.
The maximum reaction time of the device on error is less than
10 minutes.
Use of the device in a safety-related system is only permitted
within the first 20 years after production of the device. This is
a basis for the calculated failure rates.
The ambient temperature for use in a safety-related system
must be above –40 °C. The information according to the
operating instructions applies to the upper limit of the
ambient temperature.
When using the device for the measurement of corrosive
media, keep in mind the limitations referred to in the
operating instructions or data sheet.
Selection of the correct version of probe or transmitter is the
operator’s responsibility.