Mounting
26 FSM4000 D184B140U02
4.5 Information on 3A conformity
Important
Onl
y
process connections which compl
y
with "3A Sanitar
y
Standard 63" ma
y
be used.
The device must only be installed vertically with the terminal box or transmitter housing pointing
upward. The "bracket mounting (1)" option no longer applies.
G00528-01
1
Fig. 12
If concentric reducers are installed on the device, it must be mounted in a vertical position.
Refer to the section 4.6.9 "Installation in pipelines with larger nominal diameter".
G00991
Fig. 13
Please ensure that the leakage hole of the process connection is located at the deepest point of
the installed device.
G00169
1
Fig. 14
1 Leaka
g
e hole