Parameterization
70 FSM4000 D184B140U02
Submenu/parameter Value range Comment
1st line
Q [percent]; Q [unit];
Iout [mA]; Q [m/s]; Q
BargraphTotalizer;
Totalizer V;
Totalizer R;
Totalizer Diff.; HART
TAG Empty pipe
Blank
Add. choice at
enhanced diagnostic
functions:
Fprt1; Fprt2; Fprt3;
Fprt4; Hist Max Err;
Hist Min Err;
Act Max Err;
Ac Min Err;
Act Max Warn;
Act Min Warn;
Connect Warn;
Connect Err
Display 1st line
Notes regarding the values Fprt1 ... Fprt4 (assignment of
values in the submenu Diagnostic / Fingerprint) as well as
the diagnostic error register, see chapter "7.4 Additional
information regarding use of enhanced diagnostic
functions".
Q [Prozent]
1st line
For devices with PROFIBUS PA, the following options are
available: Flowrate in percentage, engineering unit, in
m/s, Bar graph, Totalizer, Totalizer F, Totalizer R,
Totalizer diff., Detector empty pipe; additional options:
PA Addr+State. The bus address and status of cyclic
communication is displayed (STOP, CLEAR or
OPERATE).
PA Adr + State
1st line
TB VolFlow Value. Displays the value for Volume_Flow
(Transducer block Index 17).
TB VolFlow Value
1st line
TB VolFlow Status. Displays the status of Volume_Flow
(Transducer block Index 17).
TB VolFlow Status
1st line
TB Total F Value. Displays the value for Totalizer F
(Transducer block Index 102).
TB Total Status. Displays the status of the Totalizer
F (Transducer block Index 102).
TB Total R Value. Displays the value for Totalizer
R (Transducer block Index 104).
TB Total R Status. Displays the status of Totalizer
R (Transducer block Index 104).
TB Total Diff Value. Displays the value for the diff.
totalizer (Transducer block Index 106).
TB Total Diff Status. Displays the status of the diff.
totalizer (Transducer block Index 106).
TBT Value
1st line
TBT Status
1st line
FB AI OUT. Displays the OUT value of the AI block. The
decimal places are generated from the OUT_SCALE
structure. The unit displayed is UNIT_INDEX from the
OUT_SCALE structure.
FB AI OUT
1st line
FB AI Status. Displays the current mode of the particular
block and the status of the output variables (OUT.
Status). The substatus is displayed after the status, if
applicable. Example: BAD 3 means status is BAD,
substatus 3 = device failure.
For information about the numeric code, refer to the
interface documentation.
FB AI Status