Parameterization
96 FSM4000 D184B140U02
The error registers for diagnostic values are structured as follows:
ID code
_______________________________
Set Off Description
S − Signal quality
L − Magnetic linearity
B − Electrode balance
e − Electrode voltage E1
E − Electrode voltage E2
D − DAC value
I − Insulation resistance
T − Coil temperature
L − Coil resistance
R − Coil voltage
C − Coil current
↓
ID
code
Register name Content of register
HMx Hist Max Errors Error / Warning: (Max.) values overshot while transmitter is in
operation.
HMi Hist Min Errors Error / Warning: (Min.) values undershot while transmitter is in
operation.
EMx Curr Max Errors Error: (Max.) present available current overshoots.
EMi Curr Min Errors Error: (Min.) present available current undershoots.
WMx Curr Max Warn Warning: (Max.) present available current overshoots.
WMi Curr Min Warn Warning: (Min.) present available current undershoots.
The information EMx _R______B__ indicates Max. Alarm in the current values for the coil
voltage and electrode balance.
The error registers for wiring errors are structured as follows:
ID code
_______________________________
Set Off Description
M − Coil short-circuit to
ground
E −
Wiring error for
electrodes
↓
ID code Register name Content of register
ConWarn Wiring warning Current warnings for wiring
ConErr Wiring Error Current errors for wiring