GLA531 Series LGR-ICOS™ Gas Analyzer
X-Purge Type Protection System
The controller of an X-Purge protection system is located in a separate explosion-proof housing above the
analyzer enclosure (see Figure 34). An X-Purge controller receives AC power only, and the voltage input is set at
the ABB factory. Refer to the certification plate on the right side panel of the enclosure for the input voltage.
ü The circuit diagrams in Figure 35 (CDA purge) and Figure 37 (N
2
purge), and the purge flow diagrams in
Figure 36 (CDA purge) and Figure 38 (N
2
purge), illustrate the design of the pressurization system and
the power interlock of X-Purge type.
ü For electrical wiring connections, refer to the
Commissioning Instructions
, document
3KXG165002R4401.
Figure 34 Enclosure Features for X-Purge Type Protection System
WARNING!
lock override switch on the enclosure right
hand side panel is only for
servicing the analyzer. Never use this switch during normal operations!
This switch must only be used by authorized service personnel and only in the
absence of a flammable atmosphere.