Navigator 500
Sodium analyzer 8 Maintenance
62 OI/ASO550–EN Rev. G
4. If the Calibrate page is not displayed, use the /
keys to scroll to the Calibrate page:
5. Press the key (below the Select prompt).
– If one or more single-stream wet-sections are
connected, the Calibrate / Sodium 1 (2, 3, 4) page is
displayed:
– If a multi-stream wet-section is connected, the
Calibrate / Sodium Three Stream page is displayed:
6. Press the key (below the Select prompt).
The following page is displayed with the Sensor
Calibration menu selected:
7. Press the key (below the Select prompt).
The Sensor Calibration page is displayed:
Press the key to highlight the Regeneration Cycle
menu.
8. Press the key (below the Select prompt).
The Regeneration Cycle page is displayed:
Press the key (below the Yes prompt) to start the
regeneration procedure.
Calibrate
Sodium 1
Back
Select
Sodium Three Stream
Calibrate
Back
Select
Sensor 1
Sensor Calibration
Scheduled Calibration
Calibration Setup
Restore Defaults
Back
Select
Sensor Calibration
Low Calibration
High Calibration
Two-Point Cal
Regeneration Cycle
Process Calibration
Back
Select
Regeneration Cycle
No
Yes
Start Procedure