REG 316*4 1MRB520049-Uen / Rev. G ABB Switzerland Ltd
10-7
10.11.2. Updating the 316EA63 firmware
The new A/D converter unit 316EA63 allows the firmware to be
downloaded without opening the unit. If updating is necessary,
this is done in a similar fashion as updating the main processor
firmware (see Section 7.5.). After each update of the main
processor firmware, the 316EA63 firmware must also be up-
dated.
When applying the DOS HMI, updating is made by calling up the
batch file ‘loadEA63.bat’, which is listed in the HMI directory.
When applying the Windows HMC CAP2/316, the item ‘EA63
download’ in the menu ‘Options’ must be selected.