42/23-81 EN Rev. 4 System Manual SCK Gas Sampling System 145
In AUTOMATIC mode the following cleaning procedure is performed when the
Sample Gas Probe Pressure MIN error message is displayed:
Phase Description
1
Two air blasts lasting 1 second each to clean the probe filter.
2
Two air blasts lasting 1 second each to clean the filter casing.
3
One air blast lasting 3 seconds to clean the probe pipe.
This cleaning cycle is repeated up to three times. If this does not remedy the error,
error message Error 3x Cleaning Not Successful Probe 60S is displayed and the
gas sampling system switches to error mode. The probe is retracted out of the kiln.
No more cleaning processes and movements take place. See Errors on page 213
section for further details.
Cleaning parameters such as the time between cleaning cycles are delivered as
default settings with the gas sampling system (Default operating parameter settings
on page 50). You can modify them in operation to suit local requirements and
specific fuels (see Optimizing operating parameters on page 53).
Configuring cleaning