ADDENDUM
12
STATIC TRIM
Use this option if the instrument is to be staticall
calibrated usin
a known pressure.
Key Display Comments
This operation will cause a chan
e in output not cor-
respondin
to the input. Be sure the control loop is in
manual.
Select
YES
.
Select
SENSOR TRIM
, or select calibration option
with down-arrow ke
and refer to the appropriate
section. Press
when done.
Select
STATIC TRIM
.
Displa
shows the value of the pressure measured
b
the 600T EN transmitter.
Enter the value of the actual static pressure usin
the
number ke
s.
CALI-
BRATE
ENTER
OUTPUT WILL BE
AFFECTED! PROCEED?
NO
→
YES
ENTER
→
SENSOR TRIM
D-TO-A ADJUST
PV BIAS
SET OUTPUT %
ENTER
[ 3 TIMES ]
ENTER
FULL TRIM
ZERO TRIM
FACTORY TRIM
→
STATIC TRIM
ENTER
STATIC PRESSURE
value
units
PRESS ENTER
TO MODIFY
ENTER
STATIC PRESSURE
value
units
ENTER
ID TAGNAME
READY