TBN480 CONDUCTIVITY TRANSMITTER
CALIBRATION
J - 7
CALIBRATE PROCESS TEMPERATURE
This section steps through calibrating the process temperature.
Calibration complete.
Key Display Comments
CALIBRATION
COMPLETE
ENTER
ID TAGNAME
READY
Key Display Comments
This operation will cause a change in output not cor-
responding to the input. Be sure the control loop is in
manual.
Select YES.
Select PROCESS CAL.
Select TEMPERATURE.
CALI-
BRATE
OUTPUT WILL BE
AFFECTED! PROCEED?
NO
YES
ENTER
CALIBRATION
PROCESS CAL
EDIT CAL
RESET CAL
ENTER
PROCESS CAL
CONDUCTIVITY
TEMPERATURE

Table of Contents