Electromagnetic Flowmeter
WaterMaster 5 Start-up and Operation
34 IM/WM Issue 7
Installation Changed
If the sensor data stored in the transmitter memory does not
match the data of the connected sensor, the warning
'INSTALLATION CHANGED' is displayed.
The changed item(s) (transmitter or sensor) can be identified
and data copied as follows:
Tran smitter
Selecting this option copies plant and stack data from the
sensor memory to the transmitter memory and loads the
totalizer from the sensor memory.
It is used to make the following changes:
Remote or integral cartridge change
Remote Tx change
New installation
Sensor
Selecting this option copies data from the transmitter memory
to the sensor and loads the totalizer from the sensor memory.
It is used to make the following changes:
Integral backplane change
Sensor change
Integral transmitter change
Process Display (Operator Page)
When the 'Process Display' (Operator Page) is displayed,
normal operation is assumed.
To access menus at a permitted access level, press the
key to display the 'Access Level' screen – see Section 5.3,
page 35.
*Example legend only
System Startup
INSTALLATION CHANGED
Identify Changed Item
Sensor
Transmitter
*Drain Water
123.45
l/s