EDUCATIONAL MATERIAL FOR RISK MINIMISATION
Display Alarm Cause Action
RunResVol
Low
Three single
signals
The reservoir volume
is low�
Change the cassette
without delay�
No
disposable,
pump will
not run
Two-tone
alarm signal
You tried to start
the pump without a
properly connected
cassette� A cassette
must be properly
attached in order for
the pump to run�
Press STOP/START
or NEXT to stop the
alarm signal� Attach the
cassette properly and
press STOP/START to
restart the pump�
Reservoir
Volume
Empty
Two-tone
alarm signal
The reservoir volume
has reached 0�0 mL�
Press STOP/START
or NEXT to silence
the alarm� Change
to a new cassette if
necessary and reset
the reservoir volume�
LowBat Three two-
tone alarm
signals
every ve
minutes
The battery power
is low but the pump
is still operating�
Change the batteries
without delay� Press
and hold STOP/
START button to
restart the pump�
Value not
saved
No alarm The input value was
not saved, i�e�, the
key ENTER/CLEAR
was not pressed�
Press NEXT to resume
programming� Save
the value before
moving on to the next
program window or
before starting the
pump�