10
electromagnetic disturbances for the expected service life.
2) Guidance and manufacturer’s declaration - Electromagnetic Emissions and Immunity.
Table 8 - Guidance and manufacturer’s declaration - electromagnetic emissions
Harmonic emissions - IEC 61000-3-2
Voltage fluctuations/ flicker emissions - IEC 61000-3-3
Table 9 - Guidance and manufacturer’s declaration - Electromagnetic Immunity
IEC 60601-1-2 - Test level
Electrostatic discharge (ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Electrical fast transient/burst
IEC 61000-4-4
Power supply lines:±2 kV
input/output lines:±1 kV
line(s) to line(s):±1 kV.; line(s) to earth:±2 kV.
100 kHz repetition frequency
Voltage dips, short
interruptions, and voltage
variations on power supply
input lines
IEC 61000-4-11
0% 0.5 cycle; At 0º, 45º, 90º, 135º, 180º, 225º, 270º
and 315º
0% 1 cycle; And 70% 25/30 cycles
Single phase: at 0; 0% 300 cycles
Power frequency magnetic field
IEC 61000-4-8
150KHz to 80MHz: 3Vrms, 6Vrms (in ISM and
amateur radio bands); 80% Am at 1kHz
Radiated RF - IEC61000-4-3
10 V/m; 80 MHz – 2,7 GHz; 80 % AM at 1 kHz
10 V/m; 80 MHz – 2,7 GHz; 80 % AM at 1 kHz
Table 10 - Guidance and manufacturer’s declaration - Electromagnetic Immunity
Radiated RF
IEC61000-4-3
(Test
specifications
for
ENCLOSURE
PORT
IMMUNITY to
RF wireless
communications
equipment)
FM
±5kHz deviation 1 kHz sine
GSM 800/900, TETRA 800,
iDEN 820, CDMA 850, LTE
Band 5
GSM 1800; CDMA 1900; GSM
1900; DECT; LTE Band1,3,4,
25; UMTS
Bluetooth, WLAN, 802.11 b/g/n,