EasyManua.ls Logo

Agilent Technologies 1260 Infinity G1315C - Page 153

Agilent Technologies 1260 Infinity G1315C
314 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 1260 Infinity DAD and MWD User Manual 153
Test Functions
8
Holmium Oxide Test
Test Failed
Probable cause Suggested actions
1
Absorbing solvent or air bubble in flow cell.
Ensure the flow cell is filled with water.
2
Incorrect calibration
Recalibrate (see “Wavelength Verification and
Calibration” on page 160) and repeat the test.
3
Dirty or contaminated flow cell.
Run the cell test (see “Cell Test” on page 156).
If the test fails, exchange the flow cell
windows.
4
Dirty or contaminated optical components
(achromat, windows).
Clean optical components with alcohol and
lint-free cloth or replace the parts (see
“Intensity Test” on page 148).
5
Old or non-Agilent lamp.
Exchange the UV lamp.

Table of Contents

Related product manuals