Agilent 1260 Infinity II Preparative Autosampler User Manual 181
Maintenance
9
Replace the Peristaltic Pump Cartridge
7
Connect the inlet filter of the solvent bottle. Use the
syringe to draw enough solvent for completely filling of
the peristaltic pump tubing before continuing to prime
the peristaltic pump.
8
Close the doors.