1260 Infinity QuatPump VL User Manual 161
Identifying Cables
11
Remote Cables
Agilent Module to 3396 Series III / 3395B Integrators
Agilent Module to Agilent 35900 A/D Converters
p/n 03396-61010 Pin 33XX Pin Agilent
module
Signal Name Active
(TTL)
9 1 - White Digital ground
NC 2 - Brown Prepare run Low
3 3 - Gray Start Low
NC 4 - Blue Shut down Low
NC 5 - Pink Not connected
NC 6 - Yellow Power on High
14 7 - Red Ready High
4 8 - Green Stop Low
NC 9 - Black Start request Low
13, 15 Not connected
p/n 5061-3378 Pin 35900 A/D Pin Agilent
module
Signal Name Active
(TTL)
1 - White 1 - White Digital ground
2 - Brown 2 - Brown Prepare run Low
3 - Gray 3 - Gray Start Low
4 - Blue 4 - Blue Shut down Low
5 - Pink 5 - Pink Not connected
6 - Yellow 6 - Yellow Power on High
7 - Red 7 - Red Ready High
8 - Green 8 - Green Stop Low
9 - Black 9 - Black Start request Low

Table of Contents