1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual 101
Error Information
6
Module Error Messages
Side Door Error
Error ID: 4750
The side door and/or the MTP board are damaged. This error message is not
displayed before the initilization is finished.
Probable cause Suggested actions
1
The side door is not installed.
Install the side door.
2
The door is bent or the magnet is
misplaced/broken.
Change the side door.
3
The sensor on the MTP board is defective.
Please contact your Agilent service
representative.
4
Excessive weight on top of the fraction
collector (see also “Initialization Failed” on
page 108 )
Check stack configuration and reduce weight
on top of the fraction collector.